PHARMA CLIMATIC TEST CHAMBERS
Series CP
For long-term stability testing with highest testing precision.
- Minimum external measurements due to compact method of construction
- Optimised parameters for highest testing precision
- Minimum noise level
Temperature range:
+10 °C / +180 °C
Humidity range:
10 – 90 % RH
Dew Point range I:
+7 °C / +45 °C
Dew Point range II:
-5 °C / +7 °C

Standard version of the product line “CP”:
- Capacitive humidity measuring system
- Casing in stainless-steel
- Door lockable
- entry port ø 50 mm
- 6 shelves
- CTS standard -terminal
- Low noise refrigerating unit
- potential free contact for malfunction signal
- RS 232 interface
CTS user benefits:
- Use of environmentally friendly materials and refrigerants
- Own CTS-controller optimised to CTS-systems
- Simple user-friendly operation
- Internal and external material in stainless-steel for easy cleaning
Temperature range:
+10 °C / +50 °C
Humidity range:
10 – 90 % RH
Dew Point range I:
+7 °C / +45 °C
Dew Point range II:
-5 °C / +7 °C
Design:
Control: microprocessor control and monitoring system
Test space: stainless steel grade 1.4301
Casing: stainless steel grade 1.4301
Refrigerants: chloride free, hermetically sealed
Humidity system: capacitive measuring
Options:
- Door with window
- Validation of the CTS unit (IQ, OQ)
- Software validation
- Test space capacity in liters: 500 / 1100
- add. entry ports ø 50 mm
- add. shelves
- Additional Pt 100 for temperature measuring on test specimen alternative reversible to controller function
- door with full-size window (reduced dew-point-range < +30 °C)
- extended temperature range down to ± 0 °C
- reservoir for humidification water
- temperature protection for test-specimen
- Touch Panel, colour display LCD
- CID software for programming and documentation
- RS 232 interface changeable to USB
Type | CP 10/500 | CP 10/1100 | ||
Test Space | liters | 500 | 1100 | |
Test Space Dimensions | height | mm, appr. | 1225 | 1225 |
width | mm, appr. | 595 | 2 x 595 = 1190 | |
depth | mm, appr. | 620 | 620 | |
Overall Dimensions | height | mm, appr. | 2080 | 2080 |
width | mm, appr. | 720 | 2 x 720 = 1440 | |
depth | mm, appr. | 850 | 850 | |
Temperature Tests | ||||
Temperature Range | °C | +10 / +50 | ||
Temperature Fluctuation | K | ± 0,5 temporally | ||
Climatic Tests | ||||
Temperature Range | °C | +10 / +50 | ||
Humidity Range | % RH | 10 / 90 | ||
Dew Point Range I | °C | +7 / +45 | ||
Dew Point Range II | °C | -5 / +7 | ||
Humidity Fluctuation | % RH | ± 1 / 3 | ||
Temperature Fluctuation | K | ± 0,5 temporally | ||
Nominal Voltage | 230 V +6% -10%, 1/N, 50 Hz | |||
Nominal Power | kW, appr. | 1,6 | 2,0 | |
Condenser | air-cooled |
All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options.
We reserve the right to make alterations due to technical development.
Click to download CTS Temperature Test Chambers PDF